Response Function of Collimated Detector for Non Axial Detector-Source Geometry

Wirawan, Rahadi and Djamal, M. and Waris, A. and Handayani, Gunawan and Kim, Hong Joo (2013) Response Function of Collimated Detector for Non Axial Detector-Source Geometry. Advanced Materials Research, 772. pp. 571-578. ISSN 1662-8985

[img] Text
AMR.772.571.pdf
Restricted to Registered users only

Download (871kB) | Request a copy
Official URL: http://doi.org/10.4028/www.scientific.net/AMR.772....

Abstract

Response function is a fundamental parameter for all detectors in order to analyze the energy distribution of gamma ray which undergoes scattering interaction with the material. The response functions of a 3 in. x 3 in. NaI(Tl) collimated detector for non axis detector-source geometry has been calculated using a Monte Carlo approach from GEANT4 simulation code with 0.662 MeV of mono-energetic of photon gamma ray. Collimated Pb with 4 cm thickness and 2 cm of holes diameter were employed for shielding. The source was assumed as an isotropic point source and it is placed at various positions to the detector axis. The comparison between the measured energy response functions and the simulated energy response functions after normalization were also performed in order to validate the modeling results.

Item Type: Article
Keywords (Kata Kunci): Response function, Collimated detector, NaI(Tl), Monte Carlo, GEANT4
Subjects: Q Science > QC Physics
Divisions: Fakultas Matematika dan ilmu Pengetahuan Alam
Depositing User: Dr. Rahadi Wirawan
Date Deposited: 13 Sep 2022 14:37
Last Modified: 13 Sep 2022 14:37